Title of article :
Ligandless-dispersive liquid–liquid microextraction of trace amount of copper ions Original Research Article
Author/Authors :
Sayed Zia Mohammadi، نويسنده , , Daryoush Afzali، نويسنده , , Yar Mohammad Baghelani، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2009
Pages :
5
From page :
173
To page :
177
Abstract :
In the present work, a new ligandless-dispersive liquid–liquid microextraction (LL-DLLME) method has been developed for preconcentration trace amounts of copper as a prior step to its determination by flame atomic absorption spectrometry. In the proposed approach 1,2-dicholorobenzene and ethanol were used as extraction and dispersive solvents, respectively. Some factors influencing on the extraction efficiency of copper and its subsequent determination were studied and optimized, such as the extraction and dispersive solvent type and volume, pH of sample solution, extraction time and salting out effect. Under the optimal conditions, the calibration curve was linear in the range of 1.0 ng mL−1–0.6 μg mL−1 of copper with R2 = 0.9985. Detection limit was 0.5 ng mL−1 in original solution (3Sb/m) and the relative standard deviation for seven replicate determination of 0.2 μg mL−1 copper was ±1.4%. The proposed method has been applied for determination of copper in standard and water samples with satisfactory results.
Keywords :
Preconcentration , Copper determination , Ligandless , Dispersive liquid–liquid microextraction
Journal title :
Analytica Chimica Acta
Serial Year :
2009
Journal title :
Analytica Chimica Acta
Record number :
1037633
Link To Document :
بازگشت