    Optimization of a modified aerospray deposition device for the preparation of samples for quantitative analysis by MALDI-TOFMS Original Research Article

  • Author/Authors

    April Holcomb، نويسنده , , Kevin G. Owens، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    7
  • From page
    49
  • To page
    55
  • Abstract
    A modified aerospray apparatus was used to prepare a thin layer sample of matrix and analyte for quantitative analysis by MALDI-TOFMS. The apparatus consists of a set of coaxial tubing; the liquid sample is forced by a syringe pump through the inner capillary and it is nebulized by a flow of gas through the outer capillary. The small droplets of sample exiting the device are deposited onto a rotating plate, which serves as the sample surface for a time-of-flight mass spectrometer. An optimization was carried out after initial experiments with the device resulted in poorer than expected reproducibility of analyte signal. A two-level plus center point factorial experiment was performed investigating several factors, including the inner capillary internal diameter, gas pressure, liquid flow, spray distance, and time. After optimization the within-sample reproducibility of the analyte signal improved 3-fold, while the sample-to-sample reproducibility improved 4.5-fold.
  • Keywords
    factorial design , Time-of-flight mass spectrometry , Optimization , Aerospray , Matrix-assisted laser desorption ionization , Sample preparation