Title of article :
Spatially addressable bead-based biosensor for rapid detection of beta-thalassemia mutations Original Research Article
Author/Authors :
Johnson Kian-Kok Ng، نويسنده , , Wen Wang، نويسنده , , Wen-Tso Liu، نويسنده , , Samuel S. Chong، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Abstract :
A simple glass-polymer bead-based biosensor was validated for the detection of beta-thalassemia mutations. Different bead types, each carrying allele-specific probes targeting a particular mutation on the beta-globin gene, were immobilized and distinguished on the chip based on their spatial addresses. Genomic DNA samples carrying various single nucleotide transitions and transversions in the beta-globin gene were subjected to polymerase chain reaction and asymmetric amplification in the presence of Cy3-labeled primers, followed by hybridization onto the chip and detection under an epifluorescent microscope. Mutations that were heterozygous or homozygous were easily detected on the device based on the signal intensity difference (or similarity) between the wildtype and mutant probes. This device successfully detected all six common beta-globin gene mutations within 30 min. The number of targeted mutations on this chip can be easily expandable through the introduction of additional probe sets.
Keywords :
Beta-globin gene , Beta-thalassemia mutations , Spatially addressable , Bead-based biosensor , Rapid detection
Journal title :
Analytica Chimica Acta
Journal title :
Analytica Chimica Acta