Title of article
An amperometric penicillin biosensor with enhanced sensitivity based on co-immobilization of carbon nanotubes, hematein, and β-lactamase on glassy carbon electrode Original Research Article
Author/Authors
Bi Chen، نويسنده , , Ming Ma، نويسنده , , Xiaoli Su، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2010
Pages
7
From page
89
To page
95
Abstract
An amperometric penicillin biosensor with enhanced sensitivity was successfully developed by co-immobilization of multi-walled carbon nanotubes (MWCNTs), hematein, and β-lactamase on glassy carbon electrode using a layer-by-layer assembly technique. Under catalysis of the immobilized enzyme, penicillin was hydrolyzed, decreasing the local pH. The pH change was monitored amperometrically with hematein as a pH-sensitive redox probe. MWCNTs were used as an electron transfer enhancer as well as an efficient immobilization matrix for the sensitivity enhancement. The effects of immobilization procedure, working potential, enzyme quantity, buffer concentration, and sample matrix were investigated. The biosensor offered a minimum detection limit of 50 nM (19 μg L−1) for penicillin V, lower than those of the conventional pH change-based biosensors by more than two orders of magnitude. The electrode-to-electrode variation of the response sensitivity was 7.0% RSD.
Keywords
Penicillin biosensor , Amperometric biosensor , Carbon nanotubes , ?-lactamase , ?-lactam antibiotics
Journal title
Analytica Chimica Acta
Serial Year
2010
Journal title
Analytica Chimica Acta
Record number
1038494
Link To Document