• Title of article

    Development of a sensitive and selective Riboflavin sensor based on carbon ionic liquid electrode Original Research Article

  • Author/Authors

    Afsaneh Safavi، نويسنده , , Norouz Maleki، نويسنده , , Hamid Ershadifar، نويسنده , , Fariba Tajabadi، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    6
  • From page
    176
  • To page
    181
  • Abstract
    The electrochemical properties of Riboflavin adsorbed on carbon ionic liquid electrode (CILE) were studied by cyclic voltammetry. A film with a surface coverage of up to 3.3 × 10−9 mol cm−2 was formed after 10 min exposure time. Electron transfer coefficient and rate constant of electron transfer across the modified electrode were found to be 0.43 and 3.03 s−1, respectively. Differential pulse voltammetry was used for the determination of Riboflavin. Two linear working ranges of 0.8–110 nM and 0.11–1.0 μM were obtained with correlation coefficients of 0.998 and 0.996, respectively. The experimental detection limit was obtained as 0.1 nM. The relative standard deviation for five replicate analyses was 4.7%. Other soluble vitamins had no significant interferences and the electrode was used for the determination of Riboflavin in pharmaceutical products, nutrition and beverages.
  • Keywords
    Electrochemical determination , Carbon ionic liquid electrode , Riboflavin , Sensor
  • Journal title
    Analytica Chimica Acta
  • Serial Year
    2010
  • Journal title
    Analytica Chimica Acta
  • Record number

    1038503