• Title of article

    Design-optimization and material selection for a femoral-fracture fixation-plate implant

  • Author/Authors

    M. Grujicic، نويسنده , , G. Arakere، نويسنده , , X. Xie، نويسنده , , M. LaBerge، نويسنده , , A. Grujicic، نويسنده , , D.W. Wagner، نويسنده , , A. Vallejo، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2010
  • Pages
    11
  • From page
    3463
  • To page
    3473
  • Abstract
    The problem of size/thickness optimization of a distal femoral-fracture fixation-plate is addressed computationally using a combined finite-element/design-optimization procedure. To obtain realistic physiological loading conditions associated with normal living activities (cycling, in the present case), a musculoskeletal multi-body inverse-dynamics analysis is carried out of a human riding the bicycle. While optimizing the design of the femoral-fracture locking-plate, realistic functional requirements pertaining to attain the required level of fracture-femur fixation and longevity/lifecycle were used. It is argued that these types of analysis should be used to complement pre-clinical implant-evaluation tests, the tests which normally include a limited number of physiological loading conditions and single pass/fail outcomes/decisions with respect to a set of lower-bound implant–performance criteria.
  • Keywords
    Musculoskeletal modeling and simulations , Optimization , Femoral-fracture fixed-plate implant
  • Journal title
    Materials and Design
  • Serial Year
    2010
  • Journal title
    Materials and Design
  • Record number

    1069042