Title of article :
Blood flow velocity and ultra-filtration velocity measured by CT imaging system inside a densely bundled hollow fiber dialyzer
Author/Authors :
Junfeng Lu، نويسنده , , Wen-Qiang Lu، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Pages :
7
From page :
1844
To page :
1850
Abstract :
Hollow fiber dialyzer is a device used for years in the therapy of hemodialysis treating people with severe kidney problems. In the design of such a device, the measurement of the blood flow velocity and the ultra-filtration velocity inside the dialyzer are very important. And non-invasive measurement solutions are highly desired during experiment. This paper presents a novel non-invasive measurement method to handle this work. The method adopts a CT (Computed Tomography) system to acquire blood motion data inside a densely bundled hollow fiber dialyzer, and then, by adopting a series of image processing methods, the blood flow velocity as well as the ultra-filtration velocity is measured.
Keywords :
Ultra-filtration velocity measurement , Densely bundled hollow fiber dialyzer , Dialyzer design , CT imaging technology , Blood flow velocity measurement