• Title of article

    A ‘3-body’ abrasion wear study of bioceramics for total hip joint replacements

  • Author/Authors

    L. Ma، نويسنده , , W.M. Rainforth، نويسنده , , D. Sun، نويسنده , , J.A Wharton، نويسنده , , R.J.K. Wood، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2009
  • Pages
    10
  • From page
    2122
  • To page
    2131
  • Abstract
    The current study focuses on the effect of the material type and the lubricant on the abrasive wear behaviour of two important commercially available ceramic on ceramic prosthetic systems, namely, Biolox® forte and Biolox® delta (CeramTec AG, Germany). A standard microabrasion wear apparatus was used to produce ‘3-body’ abrasive wear scars with three different lubricants: ultrapure water, 25 vol% new-born calf serum solution and 1 wt% carboxymethyl cellulose sodium salt (CMC-Na) solution. 1 μm alumina particles were used as the abrasive. The morphology of the wear scar was examined in detail using Atomic Force Microscopy (AFM) and Scanning Electron Microscopy (SEM). Subsurface damage accumulation was investigated by Focused Ion Beam (FIB) cross-sectional milling and Transmission Electron Microscopy (TEM). The effect of the lubricant on the ‘3-body’ abrasive wear mechanisms is discussed and the effect of material properties compared.
  • Keywords
    Bioceramics , Microabrasion , Protein adsorption , Subsurface damage
  • Journal title
    Wear
  • Serial Year
    2009
  • Journal title
    Wear
  • Record number

    1091381