Title of article :
Helium liquefaction with a commercial 4 K Gifford-McMahon cryocooler Original Research Article
Author/Authors :
P. Schmidt-Wellenburg، نويسنده , , O. Zimmer، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
5
From page :
799
To page :
803
Abstract :
This paper describes helium liquefaction using a commercial cryocooler with 1.5 W cooling power at 4.2 K (Sumitomo model RDK415D with compressor CSW-71D, consuming 6.5 kW electrical power), equipped with heat exchangers for precooling the incoming gas. No additional cooling power of cryoliquids or additional Joule–Thomson stages were utilized. Measurements of the pressure dependence of the liquefaction rate were performed. A maximum value of 83.9 g/h was obtained for 2.25 bar stabilized input pressure. Including the time needed to cool the liquefied helium to 4.2 K at 1 bar after filling the bottle connected to the cold head, and correcting for heat screen influences, this results in a net liquefaction rate of 67.7 g/h. Maintaining a pressure close to 1 bar above the bath during liquefaction, a rate of 55.7 g/h was obtained. The simple design enables many applications of the apparatus.
Keywords :
Gifford-McMahon (E) , Heat exchanger (E) , Pulse tube (E) , Cryostats (F) , Normal helium (B)
Journal title :
Cryogenics
Serial Year :
2006
Journal title :
Cryogenics
Record number :
1172706
Link To Document :
بازگشت