• Title of article

    A vibration free cryostat using pulse tube cryocooler Original Research Article

  • Author/Authors

    Chao Wang، نويسنده , , John G. Hartnett، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    6
  • From page
    336
  • To page
    341
  • Abstract
    This paper introduces a new vibration free cryostat cooled by liquid helium and a 4 K pulse tube cryocooler. The cryogenic device mounts on the sample cooling station which is cooled by liquid helium. The boil off helium is recondensed by the pulse tube cryocooler, thus the cryostat maintains zero boil off. There is no mechanical contact between the cryogenic part of the cryocooler and the sample cooling station. A bellows is used to isolate the vibration which could transfer from the cryocooler flange to the cryostat flange at the room temperature. Any vibrations generated by the operation of the cryocooler are almost entirely isolated from the cryogenic device. The cryostat provides a cooling capacity of 0.65 W at 4.21 K on the sample cooling station while maintaining a vapor pressure of 102 kPa. The sample cooling station has a very stable temperature with oscillations of less than ±3 mK during all the operations. A cryogenic microwave oscillator has been successfully cooled and operated with the cryostat.
  • Keywords
    F. Cryostat , C. Vibration free , E. Pulse tube cooler , B. Liquid helium
  • Journal title
    Cryogenics
  • Serial Year
    2010
  • Journal title
    Cryogenics
  • Record number

    1173055