Title of article :
Development of a quartz crystal microbalance (QCM) immunosensor for the detection of Listeria monocytogenes
Author/Authors :
R.D Vaughan، نويسنده , , C.K O’Sullivan، نويسنده , , G.G. Guilbault، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
4
From page :
635
To page :
638
Abstract :
This study presents the development of a QCM immunosensor for the detection of Listeria monocytogenes. A self-assembled monolayer (SAM) of thiosalicylic acid is incorporated for the covalent attachment of antibodies to the gold surface of the piezoelectric crystal. A non-Sauerbrey increase in frequency is observed upon exposure of such a crystal to specific antigen cells. This unexpected response is consistent with the obtained results and is shown to be specific. The sensor can detect L. monocytogenes cells in real time in solution to 1 × 107 cells/ml. The sensor is reusable more than 10 times without detectable loss in activity and shows negligible response to a non-specific pathogen, Bacillus cereus. The lifetime of the thiolated crystal was also investigated.
Keywords :
Listeria monocytogenes , Quartz Crystal Microbalance , Self-assembled monolayer
Journal title :
Enzyme and Microbial Technology
Serial Year :
2001
Journal title :
Enzyme and Microbial Technology
Record number :
1173532
Link To Document :
بازگشت