Title of article
Applicability of 8OCB for temperature calibration of temperature modulated calorimeters
Author/Authors
Edgar C. Schick، نويسنده , , U. JOnsson، نويسنده , , T. Vassiliev، نويسنده , , A. Minakov، نويسنده , , J. Schawe، نويسنده , , R. Scherrenberg، نويسنده , , D. Lorinczy، نويسنده ,
Issue Information
دوهفته نامه با شماره پیاپی سال 2000
Pages
9
From page
53
To page
61
Abstract
Calibration of temperature modulated calorimeters (TMC) is essential because it is often not possible to calculate heat capacity and sample temperature directly from the quantities measured. The question arises whether temperature calibration used in conventional calorimetry is also valid for TMC’s. To prove this, a well defined transition is needed which does not disturb the temperature profile within the sample during modulation. The response of the whole system (calorimeter, sample, transition) must be linear. Some low energy liquid crystal transitions behave in this way, under certain conditions, and can be used for temperature calibration in TMC’s. The procedure is demonstrated for the nematic to smectic-A transition in a cyanobiphenyl liquid crystal (8OCB). The applicability of 8OCB for temperature calibration is shown for temperature modulated DSC, AC calorimeter, 3ω method and photo-acoustic method.
Keywords
Liquid crystal , Complex heat capacity , Phase transition , Temperature modulated calorimetry , Temperature calibration