Title of article :
A calibration of the various heat-conduction paths for a heat-flux-type temperature-modulated DSC
Author/Authors :
I Moon، نويسنده , , R Androsch، نويسنده , , B Wunderlich، نويسنده ,
Issue Information :
دوهفته نامه با شماره پیاپی سال 2000
Pages :
7
From page :
285
To page :
291
Abstract :
A study was made of the effects of the heat-conducting paths from the temperature sensor to the sample and from the sample-temperature to the reference-temperature sensors on the determination of heat capacity in a temperature-modulated DSC (TMDSC) of the heat-flux type. These effects are usually assumed to be small. This is shown not to be so, and calibration procedures are developed. Using a simple model calculation, it is demonstrated that the phase angle difference between the sample temperature and the heat-flow rate should be taken into account when determining heat capacity. The results are experimentally verified for aluminum samples with various masses in quasi-isothermal experiments for a wide range of frequencies of temperature modulation for a Mettler–Toledo 820 DSC using no reference pan to simplify the analysis.
Keywords :
Temperature-modulated DSC , Heat capacity , Phase angle , calibration , Heat-flux DSC
Journal title :
Thermochimica Acta
Serial Year :
2000
Journal title :
Thermochimica Acta
Record number :
1194809