Title of article :
Evaluation of the calibration errors on cooling of a differential scanning calorimeter using different sets of standard metals
Author/Authors :
M.J.A. Malheiro، نويسنده , , J.A. Martins، نويسنده , , J.J.C. Cruz Pinto، نويسنده ,
Issue Information :
دوهفته نامه با شماره پیاپی سال 2004
Pages :
7
From page :
155
To page :
161
Abstract :
The temperature calibration on cooling of thermal analysis instruments is important for the accurate study of the non-isothermal crystallization kinetics of semi-crystalline polymers. In previous works, a methodology was proposed for performing the calibration on cooling of differential scanning calorimeters (DSCs) with standard metals, and the calibration errors were checked using transitions of high-purity liquid crystals. In this work, alternative, physically meaningful, procedures for carrying out the calibration on cooling are analyzed and validated. The calibration errors are evaluated also with liquid crystalline transitions, when the calibration is performed with standard metals, in a wide temperature range, and when due account is taken for the need of isothermal corrections to the temperature measurements. It is shown that any pair of standard metals may be used to calibrate on cooling, that the calibration errors increase for wider working temperature ranges and that, providing that certain conditions are fulfilled, both calibration procedures yield similar results.
Keywords :
DSC temperature calibration on cooling , True sample temperatures
Journal title :
Thermochimica Acta
Serial Year :
2004