• Title of article

    Isothermal microcalorimetry perfusion experiments: A method to verify the relative vapour pressure of organic liquids inside the sample vessel

  • Author/Authors

    Julia Baronsky، نويسنده , , Sabine Bongaerts، نويسنده , , Michael Traeubel، نويسنده , , Nora Urbanetz، نويسنده ,

  • Issue Information
    دوهفته نامه با شماره پیاپی سال 2009
  • Pages
    5
  • From page
    37
  • To page
    41
  • Abstract
    Perfusion microcalorimetry is frequently applied to expose pharmaceutical solids to specific relative vapour pressures (RVP) of organic liquids or water. The aim was to set up a general procedure to check the performance of a vapour pressure control device used to create a specific RVP. This was achieved by identifying and checking the significant factors in the creation of the RVP inside the reaction vessel. The flow switch valve was tested by volume flow measurements whereas the saturation of the wet line was verified in a calorimetric experiment employing methanol as model solvent. The results of the measurements were used to calculate the RVP inside the reaction vessel. The method developed is suitable to test the vapour pressure control device. It can be applied for any solvent.
  • Keywords
    Relative vapour pressure , Perfusion , Vapour sorption , Isothermal microcalorimetry
  • Journal title
    Thermochimica Acta
  • Serial Year
    2009
  • Journal title
    Thermochimica Acta
  • Record number