• Title of article

    TG–MS of air-sensitive compounds in argon

  • Author/Authors

    Stefan Hoffmann، نويسنده , , Marcus Schmidt، نويسنده , , Susann Scharsach، نويسنده , , Rudiger Kniep، نويسنده ,

  • Issue Information
    دوهفته نامه با شماره پیاپی سال 2012
  • Pages
    7
  • From page
    204
  • To page
    210
  • Abstract
    An experimental setup is presented, which allows simultaneous thermogravimetry–mass spectrometry (TG–MS) investigations of air-sensitive compounds in argon. The setup consists of a commercially available thermobalance with mass spectrometer and a glove box. Modifications needed for installations within the glove box are summarized. The performance with respect to e.g. the background gas level and the stability of the TG signal is given together with an additional list of limitations caused by the specific conditions (Ar, temperature range 20–1200 °C). The proper function of the new setup up to 800 °C is demonstrated by the thermal decomposition of CuCl2·2H2O.
  • Keywords
    TG–MS , CuCl2 , Thermogravimetry , Air-sensitive , mass spectrometry , Glove box
  • Journal title
    Thermochimica Acta
  • Serial Year
    2012
  • Journal title
    Thermochimica Acta
  • Record number