Title of article :
Design, performance and analysis of thermal lag of the UFS1 twin-calorimeter chip for fast scanning calorimetry using the Mettler-Toledo Flash DSC 1
Author/Authors :
Sander van Herwaarden، نويسنده , , Elina Iervolino، نويسنده , , Floor van Herwaarden، نويسنده , , Ties Wijffels، نويسنده , , Archi Leenaers، نويسنده , , Vincent Mathot، نويسنده ,
Issue Information :
دوهفته نامه با شماره پیاپی سال 2011
Pages :
7
From page :
46
To page :
52
Abstract :
This paper presents a new twin-membrane calorimeter chip for fast differential scanning calorimetry (DSC) with the Flash DSC 1 of Mettler-Toledo. The thin silicon nitride membranes enable scan rates in excess of 10 kK/s in heating and up to 4 kK/s in cooling for sample masses between 100 ng and 10 μg in the temperature range of −100 °C to 450 °C. The time constant for cooling is about 12 ms, the power resolution is typically 0.1–0.5 μW, the temperature accuracy of non-calibrated chips is typically better than ±5 K. The paper also shows measurements for the scan-rate dependent thermal lag of the device, showing an empty sensor thermal lag of about 0.2 ms, and a mass dependent thermal lag of about 0.3 ms/μg for Indium for a good thermal contact between Indium and membrane.
Keywords :
Thermal lag , Fast scanning calorimetry , Calorimeter chip
Journal title :
Thermochimica Acta
Serial Year :
2011
Journal title :
Thermochimica Acta
Record number :
1201977
Link To Document :
بازگشت