Title of article
The numerical design of a parallel plate flow chamber for investigation of endothelial cell response to shear stress
Author/Authors
B.J Chung، نويسنده , , A.M. Robertson، نويسنده , , D.G Peters، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
12
From page
535
To page
546
Abstract
Parallel plate chambers are frequently used to examine the response of biological cells to a constant wall shear stress. However, the stress can vary more than 80% across the chamber due to end effects. Earlier estimates of the magnitude of this inhomogeneity used boundary layer theory and experiments. Here, the full equations for steady, three-dimensional flow in a novel parallel plate device were solved numerically and used to identify an active test region where the shear stress is within 5% of a constant value. Endothelial cells can be confined to this region to assure a nearly uniform shear stress exposure.
Keywords
Parallel plate device , Shear stress , Endothelial cells , mechanotransduction , Active test region
Journal title
Computers and Structures
Serial Year
2003
Journal title
Computers and Structures
Record number
1209073
Link To Document