Title of article :
Critical pressure for capillary valves in a Lab-on-a-Disk: CFD and flow visualization
Author/Authors :
Goncalo Silva، نويسنده , , Nuno Leal، نويسنده , , Viriato Semiao، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2010
Pages :
10
From page :
1300
To page :
1309
Abstract :
Capillary valves are used as pressure barriers to control flow sequencing in microfluidic devices. Influence of valves height on liquid flow pattern and critical pressure are studied through flow visualization and CFD predictions (Gambit® 2.2.30 and FLUENT® 6.2.16). Both hydrophilic and hydrophobic walls are studied. Results show that the surface tension plays a major role in the liquid progress through the microchannel/valve and also in the valve filling process. Critical pressure varies linearly with the valve hydraulic diameter in the range 0.91 < Dh < 3.5 [mm] according to: P = 14.14 · Dh + 47.42 [Pa].
Keywords :
Lab-on-a-Disk , Capillary valve , critical pressure , CFD , Surface Tension , Two-phase microfluidics
Journal title :
Computers and Structures
Serial Year :
2010
Journal title :
Computers and Structures
Record number :
1210661
Link To Document :
بازگشت