• Title of article

    Gradient Coil Design in Permanent Magnet MRI Original Research Article

  • Author/Authors

    Guojun Hu، نويسنده , , Yu-hua PENG، نويسنده , , Wei ZHAO، نويسنده , , Xiao-ying TANG، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    4
  • From page
    419
  • To page
    422
  • Abstract
    Based on target field method, the article designs gradient coils for a permanent MRI magnet model. By simulating, winding and testing, the result verifies the validity and feasibility of target field method used in gradient coils design of permanent magnet MRI; moreover, shielded coils can restrain eddy current and can be designed based on target field method easily. The simulated results show that the effect of restraining eddy current is notable. So, the permanent magnetic MRI gradient coil can be designed by target field method. Gradient coils generate the linearly gradient field; and shielded coils restrain the eddy current.
  • Keywords
    target field method , gradient coil , Permanent
  • Journal title
    Journal of Iron and Steel Research
  • Serial Year
    2006
  • Journal title
    Journal of Iron and Steel Research
  • Record number