Title of article :
Particle size measurement of standard reference particle candidates with improved size measurement devices
Author/Authors :
Hideto Yoshida، نويسنده , , Hiroaki Masuda، نويسنده , , Kunihiro Fukui، نويسنده , , Yusuke Tokunaga، نويسنده , , Kaoru Takarada، نويسنده , , Tomohiro Sakurai، نويسنده , , Hideaki Matsumoto، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Abstract :
In order to prepare standard reference particles, experimental and theoretical studies have been conducted on particle size measurement of two kinds of spherical glass beads. The sheath flow- type electrical sensing zone method and the revised sedimentation balance method were used for the measurement. The data were compared with those obtained by the microscopic method with a sample size greater than 10 000. The particle size distributions obtained by use of the sheath flow- type electrical sensing zone method and microscopic method are nearly equal for the two kinds of glass beads. However, deviations of less than 5% were observed between the improved sedimentation balance and microscopic methods. The experimental data of the microscopic method are within the 95% reliability region of the sheath flow-type electrical sensing zone method.
Keywords :
Stokes diameter , liquid sedimentation method , standard reference particles , microscopic method , size measurement