A selective and sensitive as well as rapid chemiluminescence (CL) flow sensor for the determination of analgin is described. The analytical reagents involved in chemiluminescence reaction, luminol and periodate, were both immobilized on an anion-exchange column. The CL signals produced by the reaction between luminol and periodate, which were eluted from the column through water injection, were decreased in the presence of analgin. Analgin was sensed by measuring the decrement of CL intensity, and which was observed linear over the logarithm of analgin concentration range of 0.1 to 50.0 ng mL−1, and the limit of detection was 0.04 ng mL−1 (3ó). At a flow rate of 2.0 mL min−1, including sampling and washing, the detection could be performed in 0.5 min with a relative standard deviation of less than 3.0%. The proposed procedure was applied successfully in the monitoring of analgin in human urine samples without any pre-treatment process. It was found that the analgin concentration reached its maximum after being orally administrated for 4 h, and the analgin metabolism ratio in 10 h was 9.28% in the body of volunteers. The flow sensor offered reagentless procedures and remarkable stability in determination of analgin, and could be easily reused over 80 h.