Preparation of an acetylene black–dihexadecyl hydrogen phosphate composite film modified glassy carbon electrode and the application in the determination of hydroxycamptothecin in blood serum
Author/Authors
Huajie Zhang ، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2005
Pages
7
From page
43
To page
49
Abstract
A novel method for the fabrication of acetylene black–dihexadecyl hydrogen phosphate (AB–DHP) composite film modified glassy carbon electrodes (GCEs) was presented, and based on this, a sensitive electrochemical method for the determination of hydroxycamptothecin (HCPT) was developed. Compared with the poor signal at GCE, the electrochemical response of HCPT at AB–DHP composite film modified GCE was greatly improved, which was confirmed by the significant enhancement of the oxidation current. After the optimization of experimental parameters, the dependence of oxidation current on HCPT concentration was explored, which exhibited a good linear relationship in the range of 1.0 × 10−9 to 1.5 × 10−6 mol l−1. The detection limit of HCPT was also examined and a low value of 5.0 × 10−10 mol l−1 for 6 min accumulation was obtained (S/N = 3). This electrode was successfully applied to the detection of HCPT in cancer patientʹs serum samples.
Keywords
Hydroxycamptothecin , Voltammetry , Dihexadecyl hydrogen phosphate , Acetylene black , Serum