Title of article :
Towards traceable particle number concentration standard: Single charged aerosol reference (SCAR)
Author/Authors :
Yli-Ojanperن، نويسنده , , Jaakko and Mنkelن، نويسنده , , Jyrki M. and Marjamنki، نويسنده , , Marko and Rostedt، نويسنده , , Antti and Keskinen، نويسنده , , Jorma، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2010
Pages :
10
From page :
719
To page :
728
Abstract :
A concept of realizing a standard for aerosol particle number concentration was tested, based on generating singly charged aerosol particles in the size range from 10 up to 500 nm. To this end, a device named single-charged aerosol reference (SCAR) was designed, built, and tested. The device is based on electrical charging of nanoparticles and subsequent growth of the particles. With an accurate measurement of volume flow and electrical current from the singly charged particles, the number concentration can be accurately, and in the end, traceably determined. Laboratory tests have shown that the device can be used to generate a narrow (GSD<1.3) particle size distribution of singly charged particles. The device can be used for traceable calibration of instruments measuring the number concentration of the particles.
Keywords :
Singly charged , Aerosol reference , Standard , Number concentration , Calibration