Title of article :
Design and performance of an optimized electrical diffusion battery
Author/Authors :
Fierz، نويسنده , , Martin and Weimer، نويسنده , , Silke and Burtscher، نويسنده , , Heinz، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2009
Pages :
12
From page :
152
To page :
163
Abstract :
Standard aerosol instruments to measure particle size distributions in the ultrafine size range are large and heavy. We built a small electrical diffusion battery, which can be carried in a backpack and thus makes size-resolved short-term personal aerosol exposition measurements possible. The instrument was designed for maximal measurable particle size range and long maintenance intervals. The optimum number of stages for the diffusion battery was found with a Monte-Carlo simulation. To assess the instrumentʹs performance, we compared size distributions measured with the electrical diffusion battery to those obtained with standard aerosol instruments (the scanning mobility particle sizer and the fast mobility particle sizer of TSI Inc.). In general, the readings of the electrical diffusion battery and those of standard instruments agree well (to within 10–20%).
Keywords :
instrumentation
Journal title :
Journal of Aerosol Science
Serial Year :
2009
Journal title :
Journal of Aerosol Science
Record number :
1385588
Link To Document :
بازگشت