Title of article :
Validation of 14 used, re-calibrated and new TSI 3790 condensation particle counters according to the UN-ECE Regulation 83
Author/Authors :
Giechaskiel، نويسنده , , Barouch and Bergmann، نويسنده , , Alexander، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2011
Pages :
9
From page :
195
To page :
203
Abstract :
Four new, four re-calibrated and six used TSI 3790 condensation particle counters (CPC) were checked using soot particles produced by a mini CAST generator. One of the re-calibrated CPC was also checked with Emery oil (PAO 4) particles (evaporation-condensation method) and its counting efficiencies were in agreement with the manufacturer’s calibration data which were determined with Emery oil particles (electrospray method). However, the counting efficiencies with the soot particles were lower indicating the importance of the calibration material (chemical composition, morphology). Since UN-ECE regulation 83 for light duty vehicles does not specify the material to be used, the validation (check) should be conducted with the same material used by the manufacturer or the differences between the materials should be taken into account. Two used CPCs showed a ∼20% decrease in their counting efficiencies after 1.5 years of measuring engine exhaust aerosol. This drift could not be recognized by the status indicators of the instrument and could be attributed to a degradation of the CPC saturator (wick). Two CPCs also showed a non-linear response; a 7% difference was found between low (10 cm−3) and high concentrations (10,000 cm−3). A non-linear response (up to 40%) was also observed when the temperature difference between the saturator and the condenser was increased to decrease the cut-off size of the CPCs.
Keywords :
drift , linearity , Counting efficiency , CAST , Emery oil
Journal title :
Journal of Aerosol Science
Serial Year :
2011
Journal title :
Journal of Aerosol Science
Record number :
1385894
Link To Document :
بازگشت