Title of article :
An ultrafine particle monitor for size-resolved number concentration measurements in atmospheric aerosols
Author/Authors :
Hillemann، نويسنده , , L. and Zschoppe، نويسنده , , A. and Caldow، نويسنده , , R. and Sem، نويسنده , , G.J. and Wiedensohler، نويسنده , , A.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2014
Pages :
11
From page :
14
To page :
24
Abstract :
An ultrafine particle monitor (UFPM) was designed and constructed to obtain size-resolved particle number concentrations of atmospheric aerosols in air quality monitoring networks. The UFPM employs a corona-jet charger for particle charging, a DMA for electrical mobility classification, and an aerosol electrometer for particle quantification based on electrical charge. It does not require a radioactive source or an alcohol-based particle detector. A sensor model of the UFPM was derived, which relates the measured electrical current to the particle number concentration. A data inversion algorithm based on the Tikhonov-regularization was implemented. It yields the number concentration in six size classes between 20 and 720 nm from a 10-min-measurement with an upper limit of 3 × 10 6 cm − 3 . A method was developed to calculate the kernel matrix from the data of comparison measurements with the UFPM and a reference mobility particle size spectrometer such as an SMPS. In experimental comparisons, the number concentrations of laboratory and atmospheric aerosols measured by the UFPM were within a range of ±20% of the reference concentration determined by an SMPS.
Keywords :
Corona-jet charger , Particle number concentration monitor , Kernel matrix , atmospheric aerosol
Journal title :
Journal of Aerosol Science
Serial Year :
2014
Journal title :
Journal of Aerosol Science
Record number :
1386418
Link To Document :
بازگشت