Title of article :
Measurement of platelet function in dogs using a novel impedance aggregometer
Author/Authors :
Kalbantner، نويسنده , , K. and Baumgarten، نويسنده , , A. and Mischke، نويسنده , , R.، نويسنده ,
Issue Information :
فصلنامه با شماره پیاپی سال 2010
Pages :
8
From page :
144
To page :
151
Abstract :
The aim of this study was to optimise the technique and establish reference values for whole blood aggregometry in dogs using a novel multiplate analyser. Measurements were performed on the hirudin-anticoagulated blood of healthy dogs using a wide range of agonists. Optimal agonist concentrations were 10 μmol/L of adenosine diphosphate, 5 μg/mL of collagen and 1 mmol/L of arachidonic acid. Ristocetin (at 0.2 and 1 mg/mL) and thrombin receptor activating peptide (TRAP-6 at 32 and 160 μmol/L) did not consistently induce platelet aggregation. Coefficients of variance for within-run imprecision (n = 10 repetitions) varied from 5% to 18%. Measurement signals were significantly higher when analyses were performed on standard samples (hirudin-anticoagulated blood) compared to citrated blood or blood samples anticoagulated with citrate buffer, regardless of whether or not re-calcification was performed (P < 0.05). The findings indicate that the analyser is suitable for the investigation of platelet aggregation in dogs and analysis should be performed on hirudin-anticoagulated blood using optimised agonist concentrations.
Keywords :
Platelet aggregation , Impedance aggregometry , Precision , reference values , anticoagulant , dog
Journal title :
The Veterinary Journal
Serial Year :
2010
Journal title :
The Veterinary Journal
Record number :
1394282
Link To Document :
بازگشت