Title of article :
Surface detection errors cause overestimation of the modulus in nanoindentation on soft materials
Author/Authors :
Kaufman، نويسنده , , Jessica D. and Klapperich، نويسنده , , Catherine M.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2009
Pages :
6
From page :
312
To page :
317
Abstract :
The accuracy of mechanical properties from depth-sensing indentation, or nanoindentation, depends on the accuracy of the displacement measurement used to calculate these properties. Here, current nanoindentation techniques and analysis methods for accurate displacement measurements are reviewed. First, the ability of a commercial instrument to sense the surface of soft materials is examined. Second, methods of sample surface detection are reviewed. Finally, a case of overestimation of the elastic modulus of a compliant material using nanoindentation with incorrect displacement values is presented.
Journal title :
Journal of the Mechanical Behavior of Biomedical Materials
Serial Year :
2009
Journal title :
Journal of the Mechanical Behavior of Biomedical Materials
Record number :
1404370
Link To Document :
بازگشت