• Title of article

    Short-term fatigue testing can predict medium-term pericardium behaviour

  • Author/Authors

    Claramunt، نويسنده , , R. and Garcيa Pلez، نويسنده , , J.M. Traviesa-Alvarez، نويسنده , , L. and Spottorno، نويسنده , , J. and Ros، نويسنده , , A. and Casado، نويسنده , , M.C.، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2011
  • Pages
    7
  • From page
    1929
  • To page
    1935
  • Abstract
    The medium-term fatigue behaviour of calf pericardium (similar to the one used to manufacture cardiac bioprostheses valve leaflets) has been studied. ples were tested under fatigue subjecting them to biaxial stress at 1 Hz frequency for 5000 cycles, in 4 series of 24 samples, at several supra-physiological mean pressures and pressure amplitudes. Short-term damage parameters such as the accumulated energy consumption in 10 cycles ( E 10 ) and medium-term ones after 5000 cycles like total energy consumption ( E t ) and maximum displacement of the membrane ( D t ) have been evaluated. showed exponential growing tendency with pressure and linear tendency with pressure amplitude when only one parameter curve was plotted. Similar results were found when analysing E t and D t . correlation models were established between E 10 and E t and E 10 and D t . Similar results were achieved in the four series, with excellent determination coefficients. sults confirm that the fatigue behaviour from the very first cycles of the test can predict the medium-term behaviour of the tissue by means of measurement of suitable damage markers. ndencies observed between the parameters seem to show that the results could have been the same ones if the test had been performed at physiological pressures and amplitudes. ork opens the door to a non-destructive test of the tissue prior to employ it to manufacture valve leaflets.
  • Keywords
    Heart valve , mechanical testing , Fatigue , pericardium , Bioprosthesis
  • Journal title
    Journal of the Mechanical Behavior of Biomedical Materials
  • Serial Year
    2011
  • Journal title
    Journal of the Mechanical Behavior of Biomedical Materials
  • Record number

    1405045