• Title of article

    Flexural mechanical properties of porcine aortic heart valve leaflets

  • Author/Authors

    P. Ragaert، نويسنده , , Kim and De Somer، نويسنده , , Filip and Somers، نويسنده , , Pamela and De Baere، نويسنده , , Ives and Cardon، نويسنده , , Ludwig and Degrieck، نويسنده , , Joris، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2012
  • Pages
    7
  • From page
    78
  • To page
    84
  • Abstract
    Freshly excised porcine aortic heart valve cusps were subjected to a uni-axial flexural indentation test, from which the rupture characteristics and a functional stiffness parameter were determined. found that the flexural mechanical properties of aortic valve cusps (i) are unaffected by their coronary position and (ii) are sensitive to the effect of mechanical preconditioning. sulting values of the cuspʹs flexural mechanical properties are intended as a set of reference properties which scaffolds, meant for the tissue engineering of heart valves, must approximate in order to be considered as a functional replacement.
  • Keywords
    Heart valve , mechanical properties , Aorta , Cusps , flexure
  • Journal title
    Journal of the Mechanical Behavior of Biomedical Materials
  • Serial Year
    2012
  • Journal title
    Journal of the Mechanical Behavior of Biomedical Materials
  • Record number