• Title of article

    Virtual evaluation of stent graft deployment: A validated modeling and simulation study

  • Author/Authors

    De Bock، نويسنده , , S. and Iannaccone، نويسنده , , F. and De Santis، نويسنده , , G. and De Beule، نويسنده , , M. and Van Loo، نويسنده , , D. and Devos، نويسنده , , D. and Vermassen، نويسنده , , F. and Segers، نويسنده , , P. and Verhegghe، نويسنده , , B.، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2012
  • Pages
    11
  • From page
    129
  • To page
    139
  • Abstract
    The presented study details the virtual deployment of a bifurcated stent graft (Medtronic Talent) in an Abdominal Aortic Aneurysm model, using the finite element method. The entire deployment procedure is modeled, with the stent graft being crimped and bent according to the vessel geometry, and subsequently released. The finite element results are validated in vitro with placement of the device in a silicone mock aneurysm, using high resolution CT scans to evaluate the result. The presented work confirms the capability of finite element computer simulations to predict the deformed configuration after endovascular aneurysm repair (EVAR). These simulations can be used to quantify mechanical parameters, such as neck dilations, radial forces and stresses in the device, that are difficult or impossible to obtain from medical imaging.
  • Keywords
    Stent graft , Silicone validation , deployment , Finite element , Abdominal Aortic Aneurysm
  • Journal title
    Journal of the Mechanical Behavior of Biomedical Materials
  • Serial Year
    2012
  • Journal title
    Journal of the Mechanical Behavior of Biomedical Materials
  • Record number

    1405463