Title of article :
Debonding of porous coating of a threaded acetabular component: Retrieval analysis
Author/Authors :
?apaj، نويسنده , , ?ukasz and Markuszewski، نويسنده , , Jacek and Rybak، نويسنده , , Tomasz and Wierusz-Koz?owska، نويسنده , , Ma?gorzata، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2013
Abstract :
This report presents a case of debonding of plasma sprayed porous titanium coating from a threaded acetabular component which caused aseptic loosening of the implant. Weight bearing after delamination caused abrasive damage of the acetabular shell, and particles of the coating embedded in the acetabular liner. Microscopic examination of periprosthetic tissues showed presence of metal particles and macrophage infiltration. Despite microscopic examination of the retrieved component the cause of debonding remains unclear.
Keywords :
Aseptic loosening , Porous coating debonding , Fatigue Damage
Journal title :
Journal of the Mechanical Behavior of Biomedical Materials
Journal title :
Journal of the Mechanical Behavior of Biomedical Materials