• Title of article

    Tubular amperometric high-temperature sensors: simultaneous determination of oxygen, nitrogen oxides and combustible components

  • Author/Authors

    Somov، نويسنده , , S.I and Reinhardt، نويسنده , , G and Guth، نويسنده , , U and Gِpel، نويسنده , , W، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    2
  • From page
    68
  • To page
    69
  • Abstract
    Tubular amperometric sensors based on stabilized zirconia solid electrolyte with two working electrodes have been studied experimentally with respect to the simultaneous measurements of oxygen and nitrogen oxides, and of oxygen and combustible gas mixtures. Sensors with different types of working electrodes have been optimized for specific gas analytical problems. An almost complete separation of the sensor signals to defined gas components has been achieved by optimization of the working parameters of the electrodes. Cross-sensitivities to oxygen-containing and to combustible gas components have been measured. Cross-interference effects from CO2 and H2O at O2/NOx sensors were minimized by properly adjusting the electrode potentials. Cross-effects on O2/HCx-sensors have been found to be even lower. The results demonstrate a new level of performance for new applications of multi-electrode solid electrolyte amperometric sensors.
  • Keywords
    Amperometric high-temperature sensors , Oxygen , nitrogen oxides
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2000
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number

    1410767