Title of article :
Chemiluminescence flow biosensor for determination of total d-amino acid in serum with immobilized reagents
Author/Authors :
Li، نويسنده , , Baoxin and Zhang، نويسنده , , Zhujun، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2000
Abstract :
A flow-injection chemiluminescence (CL) biosensor for the determination of total d-amino acid is described in this paper. d-amino acid oxidase was immobilized onto amine-modified silica gel via glutaraldehyde activation and packed in a column. The other analytical reagents, including luminol and ferricyanide, were electrostatically co-immobilized on an anion-exchange column. d-amino acid was sensed by the CL reaction between hydrogen peroxide produced from the enzyme column and luminol and ferricyanide, which were eluted from the ion-exchange column through sodium phosphate injection. The linear range of d-alanine concentration is 1.1×10−6 to 1.1×10−4 mol/l and the detection limit was 4.5×10−7 mol/l (3σ). The proposed method has been successfully applied to the determination of d-amino acid in human serum.
Keywords :
d-amino acid , Chemiluminescence , Flow biosensor , Immobilized reagent
Journal title :
Sensors and Actuators B: Chemical
Journal title :
Sensors and Actuators B: Chemical