Title of article :
Simultaneous determination of oxygen transport characteristics of six membranes by hexagonal dissolved oxygen sensor system
Author/Authors :
Kim، نويسنده , , Tai-Jin and Jurng، نويسنده , , Tae-Hoon and Chung، نويسنده , , Ui-Hong and Hong، نويسنده , , Suk-In، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
10
From page :
11
To page :
20
Abstract :
In the present study, a hexagonal sensor system composed of a six port measurement device and six different cathodes with one common anode was developed for simultaneous determination of oxygen permeabilities or diffusivities of six different membranes. To minimize experimental variations caused by geometrical differences among six individual electrodes, the current output of each electrode was divided by individual correction factors determined by standardization with a typical membrane of fluorinated ethylene propylene (FEP). The hexagonal sensor system was then used to simultaneously determine the oxygen transport characteristics such as permeabilities or diffusivities of six different polypropylene membranes. The present study showed that a newly developed hexagonal dissolved oxygen sensor system can be used for simultaneous measurement of oxygen transport characteristics of six different membranes.
Keywords :
MEMBRANE , Permeability , Simultaneous , Oxygen , Sensor , Diffusivity
Journal title :
Sensors and Actuators B: Chemical
Serial Year :
2001
Journal title :
Sensors and Actuators B: Chemical
Record number :
1412658
Link To Document :
بازگشت