Title of article :
Development and characterization of an ELISA assay in PDMS microfluidic channels
Author/Authors :
Eteshola، نويسنده , , E and Leckband، نويسنده , , D، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Pages :
5
From page :
129
To page :
133
Abstract :
This report describes the first preliminary evaluation of a heterogeneous sandwich enzyme-linked immunoassay in a PDMS microfluidic device. The PDMS devices were fabricated using replica molding against a patterned photoresist generated by photolithographic techniques. With this experimental setup, the microfluidic sensor chip was successfully used to quantify a model analyte (sheep IgM) with sensitivity down to 17 nM. The conventional blocking cocktail used in nearly all polystyrene microtiter plate-based ELISA assays failed to block the nonspecific adsorption of the analyte and secondary antibody. This report describes the successful use of surface modification chemistries and a modified blocking cocktail to reduce background due to nonspecific adsorption and to thereby increase the sensitivity of the assay. These protocols can be extended to the detection of a variety of analytes by immunoassay in PDMS microchannels.
Keywords :
Polydimethylsiloxane (PDMS) , flow-through , microfluidic device , Immunoassay , Sensor
Journal title :
Sensors and Actuators B: Chemical
Serial Year :
2001
Journal title :
Sensors and Actuators B: Chemical
Record number :
1412689
Link To Document :
بازگشت