Title of article :
Immunosensing system for α-fetoprotein coupled with a disposable amperometric glucose oxidase sensor
Author/Authors :
Kim، نويسنده , , Eun Ju and Yanagida، نويسنده , , Yasuko and Haruyama، نويسنده , , Tetsuya and Kobatake، نويسنده , , Eiry and Aizawa، نويسنده , , Masuo، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2001
Abstract :
The immunosensing system coupled with a disposable amperometric glucose oxidase (GOD) sensor was developed for the determination of α-fetoprotein (AFP). The combination of the simple, portable, and low cost electrochemical measurement with the specific and sensitive enzyme-linked immunosorbent assay (ELISA) indicates a promising means to establish an immunosensing system. The results proved that the established immunosensing system was capable of detecting AFP in the low concentration value of ng ml−1. The assay took 20 min to be completed using an electrode and pre-coated microtiter plate with anti-AFP. The assay revealed linearity for AFP in the response range from 10−8 to 10−5 g ml−1. On the basis of these results, the immunosensing system offered excellent performance within 20 min compared with typical ELISA which took several hours.
Keywords :
?-Fetoprotein , Immunosensing system , Disposable amperometric sensor , Glucose oxidase
Journal title :
Sensors and Actuators B: Chemical
Journal title :
Sensors and Actuators B: Chemical