• Title of article

    A disposable electrode based on zeolite–polymer membranes for potentiometric titrations of ionic surfactants

  • Author/Authors

    Matysik، نويسنده , , S. and Matysik، نويسنده , , F.-M. and Einicke، نويسنده , , W.-D.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2002
  • Pages
    5
  • From page
    104
  • To page
    108
  • Abstract
    A planar all-solid-state electrode for the titrimetric analysis of ionic surfactants has been developed and characterised. The electrode is fabricated by lamination of suitable electrode structures that are sandwiched between flexible plastic foils. The zeolite–polymer membrane containing 50% (w/w) zeolite of the faujasite type is cast onto an exposed gold film surface to form the surfactant-sensitive electrode. ry fast potentiometric response of the sensor can be exploited for potentiometric titrations of surfactants. Without any further pre-treatment, reproducible and well-defined titration curves are obtained for the titrimetric analysis of anionic surfactants using a cationic surfactant and vice versa. According to the fabrication strategy applied, the developed devices have the character of disposable electrodes. The electrode assembly exhibits shelf-lives longer than 3 months under room temperature and dry conditions. The practical utility of the disposable all-solid-state electrodes is demonstrated in terms of titrimetric determinations of anionic surfactants in commercial detergents.
  • Keywords
    surfactant , Zeolite–polymer membrane , Disposable electrode , Potentiometry
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2002
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number

    1417064