Title of article :
A low hydraulic capacitance pressure sensor for integration with a micro viscosity detector
Author/Authors :
van der Heyden، نويسنده , , F.H.J. and Blom، نويسنده , , M.T. and Gardeniers، نويسنده , , J.G.E. and Chmela، نويسنده , , E. and Elwenspoek، نويسنده , , M. and Tijssen، نويسنده , , R. and van den Berg، نويسنده , , A.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2003
Pages :
8
From page :
102
To page :
109
Abstract :
A design is presented for a micromachined differential viscometer, that is suitable for integration into a planar hydrodynamic chromatography system (HDC) for polymer analysis. The viscometer consists of four equal flow restrictions and two pressure sensors, connected in a Wheatstone bridge configuration. this viscometer requires an ultra-low hydraulic capacitance differential pressure sensor, a pressure sensor is presented of which the internal volume displacement was reduced considerably compared to commercially available sensors. An optical method was used to detect membrane deflections with a magnitude of approximately 0.01 nm. This method lead to a linearity of 1% over the measured pressure range and a resolution of 2 Pa. his pressure sensor integrated into the viscosity detector, viscosity measurements were performed on the viscosity difference of ethanol and water.
Keywords :
Optical sensing , Integrated differential viscosity detection , Polymer analysis , Low hydraulic capacitance pressure sensor