Title of article :
Gold-film array-electrode for electrochemical ELISA
Author/Authors :
Kim، نويسنده , , S.D. and Chung، نويسنده , , J.W. and Kim، نويسنده , , J.T. and Krause، نويسنده , , H. and Pyun، نويسنده , , J.C.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2005
Abstract :
A gold-film array-electrode (8-channel) has been developed for application to an electrochemical ELISA system. The principle of the electrochemical ELISA is the amperometric quantification of oxidized 3,5,3′,5′-tetramethylbenzidine (TMB) which is widely used as a chromogenic substrate of horseradish peroxidase (HRP) in commercial ELISA kits. To make an 8-channel array-electrode, eight sets of three-electrode systems were integrated with the spacing of a 96-well microplate. The gold-film electrode for working- and auxiliary-electrode was made by sputtering a gold layer on a polyetherimide (PEI) strip, and the reference electrode was made by spin coating with Ag/AgCl paste. Using the gold-film electrode, the measurement parameters, such as analysis time, electrode area, measuring temperature and so on, were optimized. Using 50 array-electrodes, the electrode-to-electrode difference in sensitivity was estimated to be 8.2%. To demonstrate the practical applicability of the gold-film electrode, it was applied to commercial ELISA kits for the detection of HIV (antibody), HBV (antigen and antibody), and HCV (antibody). The life time of the gold-film electrode was estimated for use for over 500 repetitive measurements.
Keywords :
Amperometry , Gold-film electrode , Array-electrode , ELISA
Journal title :
Sensors and Actuators B: Chemical
Journal title :