• Title of article

    Chromium(III)-selective sensor based on tri-o-thymotide in PVC matrix

  • Author/Authors

    Gupta، نويسنده , , V.K. and Jain، نويسنده , , A.K. Nanda Kumar، نويسنده , , P. and Agarwal، نويسنده , , S. N. Maheshwari، نويسنده , , G.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    5
  • From page
    182
  • To page
    186
  • Abstract
    Tri-o-thymotide (I) has been used as an electroactive material in PVC (poly(vinyl chloride)) matrix for fabrication of chromium(III)-selective sensor. The membrane containing tri-o-thymotide, sodium tetraphenyl borate (NaTPB), dibutyl phthalate (DBP) and PVC in the optimum ratio 5:1:75:100 (w/w) exhibits a working concentration range of 4.0 × 10−6 to 1.0 × 10−1 M with a Nernstian slope of 20.0 ± 0.1 mV/decade of activity in the pH range of 2.8–5.1. The detection limit of this sensor is 2.0 × 10−7 M. The electrode exhibits a fast response time of 15 s, shows good selectivity towards Cr3+ over a number of mono-, bi- and trivalent cations and can also be used in partially non-aqueous medium (up to 15%, v/v) also. The assembly has been successfully used as an indicator electrode in the potentiometric titration of chromium(III) against EDTA and also to determine Cr(III) quantitatively in electroplating industry waste samples.
  • Keywords
    Chemical sensors , PVC , Chromium sensors , Ion-selective electrodes
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2006
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number

    1420931