Title of article :
Chromium(III)-selective sensor based on tri-o-thymotide in PVC matrix
Author/Authors :
Gupta، نويسنده , , V.K. and Jain، نويسنده , , A.K. Nanda Kumar، نويسنده , , P. and Agarwal، نويسنده , , S. N. Maheshwari، نويسنده , , G.، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2006
Pages :
5
From page :
182
To page :
186
Abstract :
Tri-o-thymotide (I) has been used as an electroactive material in PVC (poly(vinyl chloride)) matrix for fabrication of chromium(III)-selective sensor. The membrane containing tri-o-thymotide, sodium tetraphenyl borate (NaTPB), dibutyl phthalate (DBP) and PVC in the optimum ratio 5:1:75:100 (w/w) exhibits a working concentration range of 4.0 × 10−6 to 1.0 × 10−1 M with a Nernstian slope of 20.0 ± 0.1 mV/decade of activity in the pH range of 2.8–5.1. The detection limit of this sensor is 2.0 × 10−7 M. The electrode exhibits a fast response time of 15 s, shows good selectivity towards Cr3+ over a number of mono-, bi- and trivalent cations and can also be used in partially non-aqueous medium (up to 15%, v/v) also. The assembly has been successfully used as an indicator electrode in the potentiometric titration of chromium(III) against EDTA and also to determine Cr(III) quantitatively in electroplating industry waste samples.
Keywords :
Chemical sensors , PVC , Chromium sensors , Ion-selective electrodes
Journal title :
Sensors and Actuators B: Chemical
Serial Year :
2006
Journal title :
Sensors and Actuators B: Chemical
Record number :
1420931
Link To Document :
بازگشت