Title of article :
Experimental study to control the insertion resistance of internal medical instrument using magnetic field oscillation
Author/Authors :
Yoshinaka، نويسنده , , K. and Takashima، نويسنده , , K. and Okazaki، نويسنده , , T. and Ikeuchi، نويسنده , , K. and Washio، نويسنده , , T. and Chinzei، نويسنده , , K.، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2007
Abstract :
We present a new method to reduce the insertion resistance for internal medical instruments by means of vibration. A rotating permanent magnet generates an oscillated magnetic field and vibrates the internal medical instruments from outside of the body. We investigated the relation between the frequency of vibration and the insertion resistance with a pin on flat test system in vitro. As a result, vibration is effective to reduce the friction and it is possible to keep a smooth insertion by vibration.
Keywords :
Tissue , sliding friction , Catheter , Endoscope , Needle , Kinetic friction
Journal title :
Tribology International
Journal title :
Tribology International