Title of article :
The effect of acetabular cup position on wear of a large-diameter metal-on-metal prosthesis studied with a hip joint simulator
Author/Authors :
Saikko، نويسنده , , Vesa and Ahlroos، نويسنده , , Tiina and Revitzer، نويسنده , , Hannu and Ryti، نويسنده , , Oskari and Kuosmanen، نويسنده , , Petri، نويسنده ,
Issue Information :
ماهنامه با شماره پیاپی سال 2013
Abstract :
Clinically, malposition of the acetabular cup in large-diameter metal-on-metal prosthetic hip designs is associated with high wear, adverse reaction to metal debris and early failure. A steep angle of the cup (>60°) may lead to poor tribological performance. Large-diameter CoCr-on-CoCr prostheses were run in the HUT-4 hip joint simulator so that a steep angle was included. With a correct position, the tribological behaviour was excellent, the wear rate being 0.1 mm3/106 cycles. In the steepest position, lubrication failed and the wear rate was two orders of magnitude higher. This study stresses the importance of rigorous pre-clinical testing.
Keywords :
Edge loading , Abduction angle , Acetabular cup position , Metal-on-metal
Journal title :
Tribology International
Journal title :
Tribology International