Abstract
A Metrological Atomic Force Microscope (MAFM) has been constructed for the traceable calibration of transfer standards for scanning probe microscopy. It uses optical interferometry to generate image scales with direct traceability to the national standard of length. Three interferometers monitor the relative displacements of the AFM tip and sample in the x, y and z directions and the interferometer data is used directly to construct 3D images of sample surfaces. Traceable dimensional measurement of surface features may then be derived from the image data. This paper describes the MAFM instrument and presents a measurement uncertainty budget. Examples are given of measurements of pitch and step height on calibration transfer standards for scanning probe microscopy.
Keywords
Traceable calibration , measurement uncertainty , atomic force microscopy