Title of article
Calibration of linear encoders with sub-nanometer uncertainty using an optical-zooming laser interferometer
Author/Authors
Kajima، نويسنده , , Mariko and Minoshima، نويسنده , , Kaoru، نويسنده ,
Issue Information
فصلنامه با شماره پیاپی سال 2014
Pages
6
From page
769
To page
774
Abstract
The nonlinear errors of high-precision linear encoders were calibrated by using a nanometer-length calibrator that was based on the optical-zooming laser interferometer with an optical frequency comb. A transmission-type linear encoder and a reflection-type linear encoder were calibrated, and the cyclic nonlinear errors were evident. The magnitudes of the observed cyclic errors were 0.1 nm and 0.2 nm, respectively, and the best calibration uncertainties were 0.55 nm (k = 2). A traceable calibration service for linear encoders with the best calibration uncertainty in the sub-nanometer range has started based on this work.
Keywords
Length metrology , linear encoder , laser interferometer , Optical frequency comb , Optical zooming
Journal title
Precision Engineering
Serial Year
2014
Journal title
Precision Engineering
Record number
1430000
Link To Document