Title of article :
Enhancement of thermal uniformity for a microthermal cycler and its application for polymerase chain reaction
Author/Authors :
Hsieh، نويسنده , , Tsung-Min and Luo، نويسنده , , Ching-Hsing and Huang، نويسنده , , Fu-Chun and Wang، نويسنده , , Jung-Hao and Chien، نويسنده , , Liang-Ju and Lee، نويسنده , , Gwo-Bin، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2008
Pages :
9
From page :
848
To page :
856
Abstract :
Thermal uniformity is essentially important for micro reactors which require precise control of critical reaction temperatures. Accordingly, we report a new approach to increase the temperature uniformity inside a microthermal cycler, especially for polymerase chain reaction (PCR). It enhances the thermal uniformity in the reaction region of a PCR chip by using new array-type microheaters with active compensation (AC) units. With this approach, the edges of the microthermal cyclers which commonly have significant temperature gradients can be compensated. Significantly, the array-type microheaters provide higher uniformity than conventional block-type microheaters. Besides, experimental data from infrared (IR) images show that the percentages of the uniformity area with a thermal variation of less than 1 °C are 63.6%, 96.6% and 79.6% for three PCR operating temperatures (94, 57 and 72 °C, respectively) for the new microheaters. These values are significantly better than the conventional block-type microheaters. Finally, the performance of this proposed microthermal cycler is successfully demonstrated by amplifying a detection gene associated with Streptococcus Pneumoniae (S. Pneumoniae). The PCR efficiency of the new microthermal cycler is statistically higher than the block-type microheaters. Therefore, the proposed microthermal cycler is suitable for DNA amplification which requires a high temperature uniformity and is crucial for micro reactors with critical thermal constraints.
Keywords :
MEMS , Thermal uniformity , PCR , micro reactors
Journal title :
Sensors and Actuators B: Chemical
Serial Year :
2008
Journal title :
Sensors and Actuators B: Chemical
Record number :
1435756
Link To Document :
بازگشت