• Title of article

    Quick and reliable estimation of BOD load of beverage industrial wastewater by developing BOD biosensor

  • Author/Authors

    Dhall، نويسنده , , Purnima and Kumar، نويسنده , , Anil and Joshi، نويسنده , , Abha and Saxsena، نويسنده , , Tushya Kumar and Manoharan، نويسنده , , Angamuthu and Makhijani، نويسنده , , Santosh Dayal and Kumar، نويسنده , , Rita، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    6
  • From page
    478
  • To page
    483
  • Abstract
    An amperometric biosensor for determination of biochemical oxygen demand in wastewater has been developed to overcome the time consuming monitoring procedures. The performance and stability of the immobilized membrane have been investigated at 37 °C and pH 6.8. Immobilized microbial membranes maintain their stability and activity after intermittent use for 400 cycles when stored at 4 °C in sodium phosphate buffer pH 6.8. The response time of the BOD sensor was only 90 min, being independent of the concentration, and the lower detection limit was 1 mg/l. The obtained BOD values showed correlation with that of the conventional method for BOD determination (BOD5) with a deviation of ±10%. Moreover, the sensor exhibits good repeatability (3.39–4.45%) and reproducibility (1.85–2.25%). Software was added to upgrade this sensor and to make it a promising candidate for online monitoring.
  • Keywords
    Beverage wastewater , Immobilized microbial membrane , biochemical oxygen demand , BOD biosensor
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2008
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number