• Title of article

    pH-ChemFET-based analysis devices for the bacterial activity monitoring

  • Author/Authors

    Pourciel-Gouzy، نويسنده , , M.L. and Assié-Souleille، نويسنده , , S. and Mazenq، نويسنده , , L. and Launay، نويسنده , , J. and Temple-Boyer، نويسنده , , P.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2008
  • Pages
    6
  • From page
    339
  • To page
    344
  • Abstract
    Silicon and polymer microtechnologies have been developed in order to integrate pH-metry techniques in the frame of medical diagnosis. Thus, a fluidic analysis device has been designed and realised in order to monitor pH-related bacterial activities. It includes a SiO2/Si3N4 pH-sensitive chemical field effect transistor (pH-ChemFET), its titanium/gold pseudo-reference gate electrode and a poly-dimethylsiloxane (PDMS) integrated flow-cell (total volume: ≤2 mm3). The whole analysis device has been used to detect the biological activities of the Lactobacillus crispatus bacteria and to estimate its sensitivity to antibiotics. Results demonstrate the detection of pH-related bacterial metabolisms in microvolumes, enabling to reduce significantly the analysis response time with regards to standard procedures and to the development of pH-ChemFET-metry for medical analysis.
  • Keywords
    Packaging , Bacterial activity detection , pH-ChemFET , PDMS flow-cell microdevice
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2008
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number