Title of article :
Continuous measurement of acute toxicity in water using a solid state microrespirometer
Author/Authors :
Del Campo، نويسنده , , Fco. Javier and Ordeig، نويسنده , , Olga and Vigués، نويسنده , , Nuria and Godino، نويسنده , , Neus and Mas، نويسنده , , Jordi and Muٌoz، نويسنده , , Francesc Xavier، نويسنده ,
Issue Information :
روزنامه با شماره پیاپی سال 2007
Abstract :
This work presents the first true solid-state microrespirometer. It consists on a naturally developed biofilm of Pseudomonas aeruginosa over a Nafion modified array of gold microdisc electrodes. Such a device can be used to monitor acute toxicity in water streams based on the oxygen reduction current recorded at the microelectrode array. When a healthy biofilm is exposed to a toxic stream the bacteria conforming stop breathing which results in an immediate increase of the oxygen supply reaching the microelectrode array.
brication and use of such a microrespirometer is described, and a possible operation mode for a real application is presented. Given that different waters may contain toxic compounds of diverse nature, the respirometer presented here provides qualitative information only.
Keywords :
Microrespirometry , Linear sweep voltammetry , Biofilm , Ultramicroelectrode arrays , oxygen reduction
Journal title :
Sensors and Actuators B: Chemical
Journal title :
Sensors and Actuators B: Chemical